Dosing variability with pre-filled syringes of aflibercept presents risk of overdose

Article

Dr Roger A. Goldberg, speaking at the American Academy of Ophthalmology 2021 annual meeting in New Orleans, noted that small misalignments in the plunger can lead to possible overdosing.

Dosing variability with pre-filled syringes of aflibercept presents risk of overdose

There is a potential for overdosing of aflibercept (Eylea, Regeneron) resulting from small misalignments of the plunger when administering the drug in prefilled syringes, according to Roger A. Goldberg, MD, MBA, in private practice in Walnut Creek, California.

Goldberg presented the information at the American Academy of Ophthalmology 2021 annual meeting in New Orleans.

According to Goldberg, this can lead to a doubling of the volume delivered using these plungers, and this is likely due to the internal diameter of the syringes and the wide dose mark thickness.

This overdosing is probably the culprit in the increasing reports of intraocular pressure (IOP) elevations and transient central retinal artery occlusion since the syringes came into use in August 2019.

The prefilled syringes were designed to store and deliver aflibercept 2.0 mg/0.05 mL. The product is packaged as a 1-mL glass syringe with a domed-plunger tip and a dose mark to facilitate proper volume delivery for intravitreal injection, he described.

Goldberg performed a study of the characteristics and design of the prefilled syringes and tested the delivered dose across a range of potential plunger settings for the aflibercept prefilled syringes.

Images of plunger / dose mark alignment strategies. (a) ranibizumab PFS with plunger aligned according to IFU (condition 1) (b) ranibizumab PFS with plunger cone tip aligned to dose mark (condition 2) (c) aflibercept PFS with plunger aligned according to IFU (condition 3) (d) aflibercept PFS with leading plunger rib aligned to top (distal to needle) edge of dose mark (condition 4) (e) aflibercept PFS with plunger cone tip aligned to top (distal from needle) edge of dose mark (condition 5).

Images of plunger / dose mark alignment strategies. (a) ranibizumab PFS with plunger aligned according to IFU (condition 1) (b) ranibizumab PFS with plunger cone tip aligned to dose mark (condition 2) (c) aflibercept PFS with plunger aligned according to IFU (condition 3) (d) aflibercept PFS with leading plunger rib aligned to top (distal to needle) edge of dose mark (condition 4) (e) aflibercept PFS with plunger cone tip aligned to top (distal from needle) edge of dose mark (condition 5). (Images courtesy of Dr Roger A. Goldberg)

Using purified water, he assessed 3 potential syringe settings in 15 syringes: 1, plunger base aligned to proximal edge of dose mark (the manufacturer’s instruction for use); 2, plunger base aligned to distal edge of dose mark; and 3, plunger tip aligned to distal edge of dose mark.

“The internal diameter of the aflibercept prefilled syringe is 6.35 mm, leading to a volume delivered of 31.7 uL/mm,” Goldberg said. “Digital calliper measurement of the dose mark revealed a thickness of 0.75 mm, leading to a potential dosing error due to the dose mark thickness of 23.8 uL. Dose-delivery testing showed the respective averages ± standard deviations delivered dose across scenarios 1, 2, and 3: 54.8 ± 5.7 uL 78.9 ± 5.7 uL, and 109.7 ±3.4 uL.”

These differences reached significance (p < 0.05).

“Small plunger misalignments can lead to administration of double the volume of aflibercept delivered using the aflibercept prefilled syringe, which likely account for the increased reports of IOP elevation since its launch. Proper adherence to the IFU [instructions for use] can minimize dosing errors.”

Related Content: Additional AAO content | Retina | Inflammation & Infection

Newsletter

Join ophthalmologists across Europe—sign up for exclusive updates and innovations in surgical techniques and clinical care.

Recent Videos
Christine Curcio, PhD, of the University of Alabama at Birmingham Heersink School of Medicine, shares histology update and revised nomenclature for OCT with Sheryl Stevenson of the Eye Care Network and Ophthalmology Times
SriniVas R. Sadda, MD, FARVO, shares key points from his retina presentation at the International SPECTRALIS Symposium
Robert Sergott, MD, describes fluorescence lifetime imaging ophthalmoscopy (FLIO) and the International SPECTRALIS Symposium – And Beyond (ISS) in Heidelberg, Germany.
Rayaz Malik, MBChB, PhD, a professor of medicine at Weill Cornell Medicine - Qatar, spoke with Ophthalmology Times Europe about his presentation. It's titled "An eye on neurodegenerative diseases: Challenging the dogma" at the International SPECTRALIS Sympsoium. In conversation with Hattie Hayes, Ophthalmology Times Europe
Anat Loewenstein, MD, describes her presentation on remote imaging for age-related macular degeneration and geographic atrophy at the International SPECTRALIS Symposium, in conversation with Hattie Hayes of Ophthalmology Times Europe
Tyson Brunstetter, OD, PhD, a US Navy Aerospace Optometrist at the NASA Johnson Space Center in Houston, Texas, shares key takeaways from his keynote at the International SPECTRALIS Symposium (ISS)
Rayaz Malik, MBChB, PhD, shares his presentation, titled An eye on neurodegenerative diseases: Challenging the dogma, at this year's International SPECTRALIS Symposium
At the Retina World Congress, Siegfried Priglinger, MD, speaks about ensuring the best outcomes for preschool-aged patients
At the 2025 ASCRS meeting, Robert Ang, MD said small aperture IOLs can benefit all patients, especially those with complex corneas or who have undergone previous corneal refractive surgery
Viha Vig, MBChB graduate student at the University of Auckland, New Zealand, discusses her poster presentation on the relationship between mitochondiral disease, Alzheimer disease, and other types of dementia.
Related Content
© 2025 MJH Life Sciences

All rights reserved.